Related Experiment Video
Updated: Apr 13, 2026

An Experimental Protocol for Femtosecond NIR/UV - XUV Pump-Probe Experiments with Free-Electron Lasers
Published on: October 23, 2018
Sensitive fast electron spectrometer in adjustable triode configuration with pulsed tunable laser for research on
S Mingels1, V Porshyn1, B Bornmann1
1FB C Physics Department, University of Wuppertal, 42119 Wuppertal, Germany.
Abstract:
We have completed an ultra-high vacuum system for sensitive fast electron spectroscopy from cold cathodes in triode configuration under high electric fields E (<100 MV/m) and pulsed tunable laser illumination (3.5 ns, 10 Hz, hν = 0.5-5.9 eV, and 0.3-17 mJ). The cathodes are prepared and inserted under clean room conditions and can be precisely 3D-positioned, cooled or heated (77-400 K). Commissioning results with the upgraded system are presented. Field emission measurements with a W tip yielded an energy resolution of 14 meV at 4 eV pass energy and a precise determination of the emitter work function, size, and temperature. Photoemission spectroscopy of short electron bunches from a virgin and laser-ablated S-GaP crystal and quantum efficiency measurements revealed surface states, energy relaxation, and band structure effects. In conclusion, this novel system is ready now for the development and characterization of photo-induced field emission cathodes.
Related Concept Videos
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Atomic Emission Spectroscopy: Instrumentation
Atomic Emission Spectroscopy: Overview
Atomic Emission Spectroscopy: Lab