Gas Chromatography-Mass Spectrometry Paired with Total Vaporization Solid-Phase Microextraction as a Forensic Tool
Published on: May 25, 2021
Hydrogen as a GC/MS carrier and buffer gas for use in forensic laboratories
Chinyere N Nnaji1, Kristina C Williams1, Jonathan M Bishop1
1Department of Chemistry, University of North Texas, Denton, TX 76201, USA.
Abstract:
A custom set of ion volumes was manufactured in order to investigate the gain and byproducts using hydrogen as a buffer gas following electron ionization in a quadrupole ion trap mass spectrometer as compared with helium. Analyses of illicit drugs such as cocaine, codeine, and oxycodone, and explosives such as TNT, RDX, and HMTD with ion volume exit orifices of 1mm, 2mm, 4mm, 6mm, 8mm and 10mm were performed using GC/MS. Strong similarities between hydrogen and helium spectra of illicit drugs and explosives provide evidence that hydrogen can be used effectively as a buffer gas in an ion trap mass spectrometer.
More Related Videos
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...