Radiation dose for normal organs by helical tomotherapy for lung cancer
Hsien-Chun Tseng1, Wen-Shan Liu2, Hsiao-Han Tsai3
1Department of Radiation Oncology, Chung Shan Medical University Hospital, Chung Shan Medical University, Taichung 40201,Taiwan, Republic of China; School of Medical Imaging and Radiological Sciences, Chung Shan Medical University, Taichung 40201, Taiwan, Republic of China; School of Medicine, Chung Shan Medical University, Taichung 40201, Taiwan, Republic of China.
Abstract:
This study derived a simple equation of effective dose (E) versus normal organ of patients with varying body weights undergoing lung cancer treatment of helical tomotherapy (TOMO). Five tissue-equivalent and Rando phantoms were used to simulate lung cancer patients. This study then measured E and equivalent dose of organ or tissues (DT) using thermoluminescent dosimetry (TLD-100H). The TLD-100H was calibrated using TOMO 6MV photons, then inserted into phantom positions that closely corresponded with the position of the represented organs and tissues. Both E and DT were evaluated by ICRP 103. Peripheral doses varied markedly at positions close to the tumor center. The maximum statistical and total errors were 16.7-22.3%. This analytical result indicates that E of Rando and tissue-equivalent phantoms was in the ranged of 9.44±1.70 (10kg) to 4.58±0.83 (90kg)mSv/Gy. Notably, E decreased exponentially as phantom weight increased. Peripheral doses were also evaluated by TLD as a function of distance from the tumor center. Finally, experimental results are compared with those in literature. These findings will prove useful to patients, physicians, radiologists, and the public.
Related Concept Videos
Biological Effects of Radiation
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...