Phantom evaluation of a navigation system for out-of-plane CT-guided puncture
L Moncharmont1, A Moreau-Gaudry2, M Medici2
1Service de radiologie et d'imagerie médicale, hôpital Nord, CHU de Grenoble, CS 10217, 38043 Grenoble cedex 9, France.
Objective:
The purpose of this phantom study was to assess a new real time electromagnetically-guided navigation system and compare it to standard computed tomography (CT) guidance.
Material And Methods:
A prospective, randomized, comparative study was carried out over a two-day period. Operators without prior experience on the new navigation system sequentially attempted to puncture two 6 mm-diameter targets (one attempt for each target) with out-of-plane trajectories using both the standard CT-guided method and the new navigation station (NAV method).
Results:
Intention-to-treat analysis was performed for 54 operators. Twenty-two operators out of 54 (40.7%) reached the target on first attempt with the NAV method versus none (0%) using CT-guidance (P<0.001). The median distance of the puncture from the center of the target was 3.7mm [Q1-Q3=2-6.7] using NAV versus 15 mm [10-20] using CT-guidance (P<0.001). Overall planning and puncture time were shorter using NAV: 76s [50-118] versus 214s [181-264] using CT-guidance (P<0.001).
Conclusion:
Novice operators consistently performed faster and more accurate phantom punctures with out-of-plane trajectories using the electromagnetically-guided navigation system than with the standard CT-guided method.