Long-term histological changes in 1 millimeter polytetrafluoroethylene (Gore-Tex) prosthetic arterial grafts
R M Samuels1, J K McGeachie, F J Prendergast
1Department of Anatomy and Human Biology, University of Western Australia, Nedlands.
Abstract:
Vein grafts are used extensively to repair blood vessels. However, when suitable vein segments are unavailable, alternative graft materials must be used. This study tested the suitability of 1 mm diameter polytetrafluoroethylene (PTFE) grafts for small-calibre arterial replacements. Grafts of 4 mm length were inserted microsurgically into the iliac arteries of 26 rats. The grafts were removed between 3 and 22 months postoperatively and examined with light microscopy and scanning and transmission electron microscopy (SEM:TEM). Anastomotic intimal hyperplasia was minimal, consisting of a few smooth muscle cells extending 100-200 microns onto the graft. The rest of the PTFE was covered with endothelium. In a few isolated areas, there was a subendothelial layer of smooth muscle, 2-10 cells thick. The short segments used in this study were highly successful, with an overall long-term patency rate of 80%. These grafts showed no evidence of excessive neo-intimal hyperplasia.