Related Experiment Video
Updated: Apr 13, 2026

Simultaneous Measurement of Turbulence and Particle Kinematics Using Flow Imaging Techniques
Published on: March 12, 2019
Independent validation of four-dimensional flow MR velocities and vortex ring volume using particle imaging
Johannes Töger1,2, Sebastian Bidhult1,3, Johan Revstedt4
1Department of Clinical Physiology, Lund University Hospital, Lund, Lund University, Lund, Sweden.
Purpose:
This study aimed to: (i) present and characterize a phantom setup for validation of four-dimensional (4D) flow using particle imaging velocimetry (PIV) and planar laser-induced fluorescence (PLIF); (ii) validate 4D flow velocity measurements using PIV; and (iii) validate 4D flow vortex ring volume (VV) using PLIF.
Methods:
A pulsatile pump and a tank with a 25-mm nozzle were constructed. PIV measurements (1.5 × 1.5 mm pixels, temporal resolution 10 ms) were obtained on two occasions. The 4D flow (3 × 3 × 3 mm voxels, temporal resolution 50 ms) was acquired using SENSE = 2. VV was quantified using PLIF and 4D flow.
Results:
PIV showed excellent day-to-day stability (R(2) = 0.99, bias -0.04 ± 0.72 cm/s). The 4D flow mean velocities agreed well with PIV (R(2) = 0.95, bias 0.16 ± 2.65 cm/s). Peak velocities in 4D flow were underestimated by 7-18% compared with PIV (y = 0.79x + 2.7, R(2) = 0.96, -12 ± 5%). VV showed excellent agreement between PLIF and 4D flow (R(2) = 0.99, 2.4 ± 1.5 mL).
Conclusion:
This study shows: (i) The proposed phantom enables reliable validation of 4D flow. (ii) 4D flow velocities show good agreement with PIV, but peak velocities were underestimated due to low spatial and temporal resolution. (iii) Vortex ring volume (VV) can be quantified using 4D flow.