Effect of interferents on the performance of direct-reading organic vapor monitors
Ryan F LeBouf1, Christopher C Coffey
1a Department of Health and Human Services , Centers for Disease Control and Prevention, National Institute for Occupational Safety and Health , Morgantown , WV , USA.
Unlabelled:
Direct-reading organic vapor monitors are often used to measure volatile organic compound concentrations in complex chemical gas mixtures. However, there is a paucity of data on the impact of multiple gases on monitor performance, even though it is known that monitor sensitivity may vary by chemical. This study investigated the effects of interferents on the performance of the MIRAN SapphIRe Portable Ambient Air Analyzer (SAP) and Century Portable Toxic Vapor Analyzer (TVA-1000) when sampling a specific agent of interest (cyclohexane). The TVA-1000 contained a dual detector: a photoionization detector (PID) and a flame ionization detector (FID). Three devices of each monitor were challenged with different combinations of cyclohexane and potential interferent vapors (hexane, methyl ethyl ketone, trichloroethylene, and toluene) at 21°C and 90% relative humidity (RH), an extreme environmental condition. Five replicates at four target concentrations were tested: 30, 150, 300, and 475 ppm. Multiple proportions of cyclohexane to interferent enabled the determination of the interferent effect on monitor performance. The monitor concentrations were compared to reference concentrations measured using NIOSH Method 1500. Three scenarios were investigated: no response factor, cyclohexane response factor, and weighted-mixed response factor applied. False negatives occurred more frequently for PID (21.1%), followed by FID (4.8%) and SAP (0.2%). Measurements from all monitors generally had a positive bias compared to the reference measurements. Some monitor measurements exceeded twice the reference concentrations: PID (36.8%), SAP (19.8%), and FID (6.3%). Evaluation of the 95% confidence intervals indicated that performance of all monitors varied by concentration. In addition, the performance of the PID and SAP varied by presence of an interfering compound, especially toluene and hexane for the PID and trichloroethylene for the SAP. Variability and bias associated with all these monitors preclude supplanting traditional sorbent-based tube methods for measuring volatile organic compounds (VOCs), especially for compliance monitoring.
Implications:
Industrial hygienists need to use care when using any of the three monitor detection types to measure the concentration of unknown chemical mixtures. Monitor performance is affected by the presence of interferents. Application of manufacturer recommended response factors may not adequately scale measurements to minimize monitor bias when compared to standard reference methods. Users should calibrate their monitors to a known reference method prior to use, if possible. Each of the monitors has its own limitations, which should be considered to ensure quality measurements are reported.
More Related Videos
05:48Analysis of Volatile and Oxidation Sensitive Compounds Using a Cold Inlet System and Electron Impact Mass Spectrometry
Published on: September 5, 2014
08:23Analyzing the Photo-oxidation of 2-propanol at Indoor Air Level Concentrations Using Field Asymmetric Ion Mobility Spectrometry
Published on: June 14, 2018
Related Concept Videos
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Atomic Emission Spectroscopy: Interference
Voltammetry: Factors Affecting Measurements
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...