First Clinical Evaluation of High-Pitch Dual-Source Computed Tomographic Angiography Comparing Automated Tube
Martin Beeres1, Kimberly Williams, Ralf W Bauer
1From the *Department of Diagnostic and Interventional Radiology, Clinic of the Goethe University, Frankfurt, Germany; and †Department of Diagnostic and Interventional Radiology, Cairo University Hospital, Cairo, Egypt.
Objective:
To investigate and compare the use of automated tube potential selection (ATPS) with automated tube current modulation (ATCM) in high-pitch dual-source computed tomographic angiography (CTA) for imaging the whole aorta without electrocardiogram synchronization.
Methods:
Two groups of 60 patients underwent CTA on a dual-source computed tomographic device in high-pitch mode: ATCM (with 100-kV fixed tube potential) was used in group 1 and ATPS (with the same image quality options) in group 2. For the evaluation of radiation exposure, CT dose index and dose-length product were analyzed. Contrast and image quality were assessed by 2 independent observers.
Results:
The ATPS group received a higher radiation dose than the ATCM group (P < 0.001) because in 80% of patients, the software switched to use of a 120-kV tube potential. In all cases, images of the aorta were of sufficient quality.
Conclusions:
High-pitch dual-source CTA of the aorta using ATPS is feasible in clinical routine, but is associated with higher radiation exposure than the ATCM protocol. This finding contradicts previously evaluations of ATPS based on single-source techniques.
More Related Videos
09:57Four-Dimensional Computed Tomography-Guided Valve Sizing for Transcatheter Pulmonary Valve Replacement
Published on: January 20, 2022
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...