Related Experiment Video
Updated: Apr 12, 2026

07:14
Detection and Quantification of Calcitonin Gene-Related Peptide CGRP in Human Plasma Using a Modified Enzyme-Linked Immunosorbent Assay
Published on: June 16, 2023
3.6K
Surface plasmon resonance-based immunoassay for human C-reactive protein.
S K Vashist1, E M Schneider, J H T Luong
1HSG-IMIT - Institut für Mikro- und Informationstechnik, Georges-Koehler Allee 103, 79110 Freiburg, Germany. sandeep.vashist@yahoo.com.
The Analyst
|May 13, 2015
Summary
A new surface plasmon resonance (SPR)-based immunoassay (IA) rapidly detects human C-reactive protein (CRP), a key biomarker for various disorders. This sensitive assay offers a faster, more stable method for CRP detection in clinical samples.
Area of Science:
- Biomedical Engineering
- Analytical Chemistry
- Immunosensing
Background:
- Human C-reactive protein (CRP) is a crucial biomarker for inflammatory conditions, metabolic disorders, and infections.
- Existing methods for CRP detection can be time-consuming or lack the required sensitivity and stability for rapid clinical diagnostics.
Purpose of the Study:
- To develop and validate a rapid, highly-sensitive surface plasmon resonance (SPR)-based immunoassay (IA) for detecting human C-reactive protein (CRP).
- To establish an efficient and stable immobilization strategy for antibodies on SPR chips for improved assay performance.
Main Methods:
- Functionalization of gold-coated SPR chips using EDC-activated protein A/G and APTES.
- Oriented immobilization of anti-human CRP capture antibodies onto the functionalized SPR chip.
- Assay validation using CRP standards and spiked human blood, serum, and plasma samples.
Main Results:
- The developed SPR-based IA demonstrated high sensitivity for CRP detection, with a limit of detection (LOD) of 1.2 ng mL(-1) and a limit of quantification (LOQ) of 4.6 ng mL(-1) within the 1.2-80 ng mL(-1) range.
- The assay accurately detected CRP in various human biological matrices, including whole blood, serum, and plasma, with precision comparable to established methods.
- Antibody-bound SPR chips maintained functional activity for up to 10 weeks when stored at 4 °C, indicating enhanced assay stability and reduced analysis time.
Conclusions:
- A rapid and highly-sensitive SPR-based immunoassay for human CRP detection has been successfully developed and validated.
- The simplified antibody immobilization strategy offers a significant reduction in assay development time and improves the stability of the sensing platform.
- This SPR-based IA shows great potential for efficient and reliable CRP monitoring in clinical diagnostics.