Related Experiment Video
Updated: Apr 12, 2026

Characterizing Far-infrared Laser Emissions and the Measurement of Their Frequencies
Published on: December 18, 2015
Absolute gauge block calibration using ultra-precise optical frequency synthesizer locked to a femtosecond comb
Abstract:
In this paper, we report a gauge block (GB) calibration that is traceable to the SI unit of time, the second. Four ultra-stable optical telecommunication wavelengths near 1556 nm are obtained by locking a narrow-tuning-range fiber laser to a fiber-based femtosecond frequency comb. Since the GB calibration system does not operate at this region of spectrum, the superior frequency stability of the laser is transferred to the 778 nm region by using a waveguide periodically poled lithium niobate crystal. After applying the locking scheme, the stability and accuracy of the laser become better than 8×10(-12). The frequency-doubled light is sent through 30 m optical fiber to a GB interferometer, which is installed at a different laboratory in the same building. Using this calibration scheme, a GB with a nominal length of 100 mm is calibrated with an uncertainty of ±52 nm. This uncertainty value is still comparable to or even better than other metrology laboratories for a similar block length.
Related Concept Videos
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...