128-slice dual-source CT coronary angiography with prospectively electrocardiography-triggered high-pitch spiral
Kosuke Matsubara1, Keita Sakuda2, Haruka Nunome2
1Department of Quantum Medical Technology, Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, Kanazawa, Ishikawa, Japan matsuk@mhs.mp.kanazawa-u.ac.jp.
Background:
Dual-source computed tomography (CT) enables CT coronary angiography (CTCA) with a prospectively electrocardiography (ECG)-triggered high-pitch spiral (HPS) mode.
Purpose:
To evaluate the radiation dose, image quality, and diagnostic acceptability of the HPS mode in CTCA and to compare HPS with the step-and-shoot (SAS) and low-pitch spiral (LPS) modes.
Material And Methods:
One hundred and thirty-eight patients who underwent CTCA with a 128-slice dual-source CT scanner were retrospectively included in this study. Seventeen patients (average heart rate of ≤65 beats per minute [bpm] prior to acquisition) were evaluated in the HPS mode, 88 (average heart rate of >65 and ≤80 bpm prior to acquisition) in the SAS mode, and 33 (average heart rate of >80 bpm prior to acquisition or patients with an unstable heart rhythm) in the LPS mode. Radiation dose and image noise were recorded for each patient. Diagnostic acceptability was graded using a four-point scale (1, unacceptable; 2, suboptimal; 3, acceptable; 4, fully acceptable).
Results:
The effective dose in the HPS mode was 1.5 ± 0.2 mSv, which was lower than that in SAS (8.9 ± 2.7 mSv) and LPS (21.5 ± 4.3 mSv) modes. There were no significant differences in the image noise levels in the descending aorta and left atrium. The average per-patient diagnostic acceptability was 3.2, 3.6, and 3.7 in HPS, SAS, and LPS modes, respectively.
Conclusion:
The radiation dose is lower with HPS than with other modes, and the HPS mode-acquired images of patients with heart rates of ≤65 bpm are nearly acceptable for diagnostic image interpretation.
More Related Videos
11:13Quantification of Mouse Heart Left Ventricular Function, Myocardial Strain, and Hemodynamic Forces by Cardiovascular Magnetic Resonance Imaging
Published on: May 24, 2021
05:32Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph
Published on: February 21, 2025
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT