Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Vector analysis of 1-year astigmatic outcomes from a prospective, randomized, fellow eye comparison of
Purpose:
To compare the astigmatic outcomes of LASIK with a single excimer laser platform using either wavefront-guided (WFG) or wavefront-optimized (WFO) modes.
Methods:
Sixty-eight eyes of 34 patients underwent LASIK for myopia with the Wavelight Allegretto Eye-Q 400-Hz laser platform (Alcon Laboratories, Inc., Hüenberg, Switzerland). One eye underwent WFG-LASIK and the contralateral eye underwent WFO-LASIK. Alpins vector analysis of astigmatism was performed using manifest refraction measured preoperatively and 12 months postoperatively and vertexed to the corneal plane. Eyes were stratified for subgroup analysis based on preoperative manifest astigmatism.
Results:
WFG-LASIK and WFO-LASIK were similar with regard to surgically induced astigmatism, difference vector, magnitude of error, correction index, flattening index, and index of success. The angle of error was 80% less in the WFG group compared to the WFO group (1.92° ± 0.67º vs 9.66° ± 3.7º, P = .04).
Conclusions:
WFG-LASIK and WFO-LASIK using the Alcon WaveLight Allegretto Eye-Q 400-Hz excimer laser platform produce similar astigmatic results in myopic patients; however, the WFG mode may produce slightly more predictable astigmatic corrections.