Obstructed kidney and sepsis secondary to urethral catheter misplacement into the distal ureter
Ruairidh Lorn Hunter Crawford1, Thomas Liston1, Ai Shiang Bong1
1Department of Urology, Worthing Hospital, Worthing, West Sussex, UK.
Abstract:
An 86-year-old woman underwent routine catheter replacement in the community. The new catheter failed to drain urine. Attempts to remove the catheter failed, both by the community nurse as well as by the urology team in the hospital. A CT scan confirmed that the catheter balloon was inflated in the distal right ureter. The patient was started on antibiotics and listed for cystoscopy under general anaesthetic. The catheter was visualised entering the right ureter and the balloon punctured using a wire under image intensifier guidance. Once removed, a new catheter was inserted. Very dilated ureteric orifices were noted. Post operatively the patient required HDU support for 48 h due to sepsis and on recovery was discharged home. The key learning point in this case is to always consider catheter misplacement in the ureter if it is not draining well and the patient presents with pain.
Related Concept Videos
Acute Kidney Injury II: Pathophysiology
Acute Pyelonephritis I: Introduction
Urinary Tract Infection II: Pathophysiology
Kidney Transplant II: Surgical Procedure
Acute Pyelonephritis II: Diagnostic Studies and Management
Disorders of the Urinary System
Urinary tract infections (UTIs) are one of the most common urinary system disorders. They are caused by bacteria that enter the urethra and can spread to the bladder resulting in cystitis. Pyelonephritis is the result of a UTI that has ascended to the level of the...