Coronary calcium scores are systematically underestimated at a large chest size: A multivendor phantom study
Martin J Willemink1, Bronislaw Abramiuc2, Annemarie M den Harder1
1Department of Radiology, University Medical Center Utrecht, P.O. Box 85500, E01.132, Utrecht 3508 GA, The Netherlands.
Objective:
To evaluate the effect of chest size on coronary calcium score (CCS) as assessed with new-generation CT systems from 4 major vendors.
Methods:
An anthropomorphic, small-sized (300 × 200 mm) chest phantom containing 100 small calcifications (diameters, 0.5-2.0 mm) was evaluated with and without an extension ring on state-of-the-art CT systems from 4 vendors. The extension ring was used to mimic a patient with a large chest size (400 × 300 mm). Image acquisition was repeated 5 times with small translations and/or rotations. Routine clinical acquisition and reconstruction protocols for small and large patients were used. CCS was quantified as Agatston and mass scores with vendor software.
Results:
The small-sized phantom resulted in median (interquartiles) Agatston scores of 10 (9-35), 136 (123-146), 34 (30-37), and 87 (85-89) for Philips, GE, Siemens, and Toshiba, respectively. Mass scores were 4 mg (3-9 mg), 23 mg (21-27 mg), 8 mg (8-9 mg), and 20 mg (20-20 mg), respectively. Adding the extension ring resulted in reduced Agatston scores for all vendors (17%-48%) and mass scores for 2 vendors (11%-49%). Median Agatston scores decreased to 9 (5-10), 79 (60-80), 27 (24-32), and 45 (29-53) units, and median mass scores remained similar for Philips at 4 mg (4-6 mg) and Siemens at 8 mg (7-8 mg) and decreased for the other vendors to 13 mg (11-14 mg) and 10 mg (8-13 mg), respectively.
Conclusion:
This multivendor phantom study showed that CCS can be underestimated up to 50% (49%-66%) for Agatston scores and 49% (36%-59%) for mass scores at a larger chest size, which may be relevant for women and large patients. However, CCS underestimation by chest size differs considerably by vendor.
More Related Videos
06:57Author Spotlight: Advancing Cardiovascular Imaging - Introducing the Spatially Weighted Calcium Score for Early Disease Detection
Published on: September 22, 2023
08:02Author Spotlight: Enhanced Quantification of Cardiovascular Calcification Progression for Longitudinal Micro PET/CT Studies in Small Research Animals
Published on: November 15, 2024
Related Concept Videos
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Acute Coronary Syndrome III: Diagnostic Studies
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...