Optimization of insulin pump therapy based on high order run-to-run control scheme
Jianyong Tuo1, Huiling Sun1, Dong Shen1
1College of Information Science and Technology, Beijing University of Chemical Technology, Beijing 100029, PR China.
Background And Objectives:
Continuous subcutaneous insulin infusion (CSII) pump is widely considered a convenience and promising way for type 1 diabetes mellitus (T1DM) subjects, who need exogenous insulin infusion. In the standard insulin pump therapy, there are two modes for insulin infusion: basal and bolus insulin. The basal-bolus therapy should be individualized and optimized in order to keep one subject's blood glucose (BG) level within the normal range; however, the optimization procedure is troublesome and it perturb the patients a lot. Therefore, an automatic adjustment method is needed to reduce the burden of the patients, and run-to-run (R2R) control algorithm can be used to handle this significant task.
Methods:
In this study, two kinds of high order R2R control methods are presented to adjust the basal and bolus insulin simultaneously. For clarity, a second order R2R control algorithm is first derived and studied. Furthermore, considering the differences between weekdays and weekends, a seventh order R2R control algorithm is also proposed and tested.
Results:
In order to simulate real situation, the proposed method has been tested with uncertainties on measurement noise, drifts, meal size, meal time and snack. The proposed method can converge even when there are ±60 min random variations in meal timing or ±50% random variations in meal size.
Conclusions:
According to the robustness analysis, one can see that the proposed high order R2R has excellent robustness and could be a promising candidate to optimize insulin pump therapy.
More Related Videos
09:08Three-dimensional Printing of Thermoplastic Materials to Create Automated Syringe Pumps with Feedback Control for Microfluidic Applications
Published on: August 30, 2018
11:49Quantitative and Temporal Control of Oxygen Microenvironment at the Single Islet Level
Published on: November 17, 2013
Related Concept Videos
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Insulin Formulations: Types and Delivery
Short-acting insulins are divided into...
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
PI Controller: Design
Time-Domain Interpretation of PD Control
Consider the example of control of motor torque. Initially, a positive...