Evolution of mammographic image quality in the state of Rio de Janeiro
Vanessa Cristina Felippe Lopes Villar1, Marismary Horsth De Seta2, Carla Lourenço Tavares de Andrade2
1Master, Health Management Analyst at Instituto Nacional de Infectologia Evandro Chagas/Fundação Oswaldo Cruz, Rio de Janeiro, RJ, Brazil.
Objective:
To evaluate the evolution of mammographic image quality in the state of Rio de Janeiro on the basis of parameters measured and analyzed during health surveillance inspections in the period from 2006 to 2011.
Materials And Methods:
Descriptive study analyzing parameters connected with imaging quality of 52 mammography apparatuses inspected at least twice with a one-year interval.
Results:
Amongst the 16 analyzed parameters, 7 presented more than 70% of conformity, namely: compression paddle pressure intensity (85.1%), films development (72.7%), film response (72.7%), low contrast fine detail (92.2%), tumor mass visualization (76.5%), absence of image artifacts (94.1%), mammography-specific developers availability (88.2%). On the other hand, relevant parameters were below 50% conformity, namely: monthly image quality control testing (28.8%) and high contrast details with respect to microcalcifications visualization (47.1%).
Conclusion:
The analysis revealed critical situations in terms of compliance with the health surveillance standards. Priority should be given to those mammography apparatuses that remained non-compliant at the second inspection performed within the one-year interval.