Related Experiment Video
Updated: Apr 12, 2026

Ultrasonography of the Adult Male Urinary Tract for Urinary Functional Testing
Published on: August 14, 2019
Comparison between uroflowmetry and sonouroflowmetry in recording of urinary flow in healthy men
Jan Krhut1,2, Marcel Gärtner3, Radek Sýkora1
1Department of Urology, University Hospital, Ostrava, Czech Republic.
Objectives:
To evaluate the accuracy of sonouroflowmetry in recording urinary flow parameters and voided volume.
Methods:
A total of 25 healthy male volunteers (age 18-63 years) were included in the study. All participants were asked to carry out uroflowmetry synchronous with recording of the sound generated by the urine stream hitting the water level in the urine collection receptacle, using a dedicated cell phone. From 188 recordings, 34 were excluded, because of voided volume <150 mL or technical problems during recording. Sonouroflowmetry recording was visualized in a form of a trace, representing sound intensity over time. Subsequently, the matching datasets of uroflowmetry and sonouroflowmetry were compared with respect to flow time, voided volume, maximum flow rate and average flow rate. Pearson's correlation coefficient was used to compare parameters recorded by uroflowmetry with those calculated based on sonouroflowmetry recordings.
Results:
The flow pattern recorded by sonouroflowmetry showed a good correlation with the uroflowmetry trace. A strong correlation (Pearson's correlation coefficient 0.87) was documented between uroflowmetry-recorded flow time and duration of the sound signal recorded with sonouroflowmetry. A moderate correlation was observed in voided volume (Pearson's correlation coefficient 0.68) and average flow rate (Pearson's correlation coefficient 0.57). A weak correlation (Pearson's correlation coefficient 0.38) between maximum flow rate recorded using uroflowmetry and sonouroflowmetry-recorded peak sound intensity was documented.
Conclusions:
The present study shows that the basic concept utilizing sound analysis for estimation of urinary flow parameters and voided volume is valid. However, further development of this technology and standardization of recording algorithm are required.
More Related Videos
06:46Author Spotlight: Enhanced Urodynamic Method for Precise Urine Measurement in Awake Mice with Neurogenic Bladder
Published on: June 7, 2024
10:07Cystometric and External Urethral Sphincter Measurements in Awake Rats with Implanted Catheter and Electrodes Allowing for Repeated Measurements
Published on: January 30, 2018
Related Concept Videos
Urodynamic Studies: Uroflowmetry
Imaging Studies VI: Voiding Cystourethrography and Cystography
Imaging Studies II: Ultrasonography
Imaging Studies V: Intravenous Urography and Retrograde Pyelography
Physiology of the Genitourinary System I: Renal Blood Flow and Glomerular Filtration
Pipe Flowrate Measurement
The orifice meter is a simple,...