Myocardial blood flow quantification by Rb-82 cardiac PET/CT: A detailed reproducibility study between two
Vincent Dunet1, Ran Klein2, Gilles Allenbach1
1Department of Nuclear Medicine and Molecular Imaging, Lausanne University Hospital, Rue du Bugnon 46, 1011, Lausanne, Switzerland.
Summary
This study compared cardiac PET software for myocardial blood flow (MBF) and myocardial flow reserve (MFR) quantification. Results show excellent concordance and precision between PMOD and FlowQuant, suggesting they are interchangeable for clinical use.
Area of Science:
- Nuclear Medicine
- Cardiovascular Imaging
- Quantitative Analysis
Background:
- Accurate myocardial blood flow (MBF) quantification from cardiac PET is crucial for clinical applications.
- Existing software packages for MBF analysis lack comparative concordance studies.
- Reproducibility of MBF measurements is essential for its full clinical potential.
Purpose of the Study:
- To compare the concordance, precision, and bias of two analysis software packages (PMOD and FlowQuant) for myocardial blood flow (MBF) and myocardial flow reserve (MFR) quantification.
- To assess the interchangeability of these software packages in clinical practice.
Main Methods:
- Dynamic Rubidium-82 (Rb-82) PET scans were performed on 51 patients at rest and during adenosine stress.
- MBF and MFR were quantified using PMOD and FlowQuant with the Lortie model and analyzed via a 17-segment model.
- Concordance analysis, including Pearson's correlation (ρ) and Lin's concordance correlation coefficient (ρc), was used to assess precision and bias.
Main Results:
- Excellent precision (ρ=0.97, ρc=0.96) and minimal systematic bias (Cb=0.99) were found for MBF quantification between PMOD and FlowQuant.
- Good precision (ρ=0.83, ρc=0.76) and a slight systematic bias (Cb=0.92) were observed for MFR.
- While PMOD showed slightly higher values at higher MBF/MFR, Bland-Altman plots indicated no significant mean bias on a per-segment basis.
Conclusions:
- Excellent concordance was demonstrated between PMOD and FlowQuant for both MBF and MFR quantification in Rb-82 cardiac PET.
- Despite minor differences at higher values, both software packages can be used interchangeably in routine clinical practice.
- These findings support the reliable use of either software for quantitative cardiac PET analysis.