Related Experiment Video
Updated: Apr 12, 2026

Author Spotlight: Technologies and Challenges in Elemental Analysis of Food Samples
Published on: December 22, 2023
[Determination of 8 elements in food by inductively coupled plasma optical emission spectrometer simultaneously]
Objective:
To establish a method for determination of 8 elements including K, Na, Ca, Mg, Fe, Zn, Cu and Mn in food by inductively coupled plasma optical emission spectrometer (ICP-OES).
Methods:
After high-pressure digestion by HNO3 in closed-vessel, 8 elements were simultaneously determined by ICP-OES.
Results:
The detection limits of 8 elements ranged from 0. 02 to 6 mg/kg(solid samples), 0. 005 mg/L to 2 mg/L(liquid samples). The average correlation coefficients were; ≥0. 999 with good linearity. The precision ranged from 0. 1% to 9. 6%, the recoveries ranged from 88% to 115%. The results of analyzing standard materials were in the certificate values.
Conclusion:
The method is rapid and accurate with less interference which meets the technical requirements of food testing and can be applied to the determination of 8 elements in food simultaneously.
More Related Videos
Related Concept Videos
Atomic Emission Spectroscopy: Lab
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Atomic Emission Spectroscopy: Instrumentation
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Atomic Emission Spectroscopy: Overview
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview