Slănic-Prahova low background calibration facility
Aurelia Celarel1, O G Duliu2, S Bercea3
1Horia Hulubei National Institute for Physics and Nuclear Engineering, 30, Reactorului Str., PO Box MG-6, Magurele (Ilfov) 077125, Romania Doctoral School on Physics, University of Bucharest, 405, Atomistilor Str., PO Box MG-11, Magurele (Ilfov) 077125, Romania.
Abstract:
The reduced background of 2 nSv h(-1) of the Slănic-Prahova Low Background Radiation Laboratory allowed installing a calibration stand for low-dose-rates dosimetry. The stand is provided with (60)Co, (137)Cs and (241)Am low activity sources. A Physikalisch-Technische Bundesanstalt-calibrated AUTOMESS 6150 AD-6 dose rate meter with a 6150 AD-b/H external probe was used to check to what extent this stand could serve as a low background calibration facility. A detailed analysis of possible uncertainties in measuring dose rates evidenced an extended uncertainty related to the certified calibration as well as instrument readings of about 3 % for a confidence level of 95 %. In these conditions, the experimentally determined dose rates for all three gamma ray sources and for source-to-probe distances varying between 0.3 and 2.0 m confirmed a good correlation between the calculated and measured dose rates. At the same time, dose rates perfectly obey to an inverse squared distances law.
More Related Videos
10:22Split Point Analysis and Uncertainty Quantification of Thermal-Optical Organic/Elemental Carbon Measurements
Published on: September 7, 2019
12:51Simultaneous Multicolor Imaging of Biological Structures with Fluorescence Photoactivation Localization Microscopy
Published on: December 9, 2013
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Calibration Curves: Correlation Coefficient