Related Experiment Video
Updated: Apr 12, 2026

Photodynamic Therapy with Blended Conducting Polymer/Fullerene Nanoparticle Photosensitizers
Published on: October 28, 2015
In-vivo light dosimetry for HPPH-mediated pleural PDT
Andreea Dimofte1, Timothy C Zhu1, Jarod C Finlay1
1Department of Radiation Oncology, University of Pennsylvania, Philadelphia, PA.
Abstract:
This study examines the light fluence (rate) delivered to patients undergoing pleural PDT as a function of treatment time, treatment volume and surface area. The accuracy of treatment delivery is analyzed as a function of the calibration accuracies of each isotropic detector and the calibration integrating sphere. The patients studied here are enrolled in a Phase I clinical trial of HPPH-mediated PDT for the treatment of non-small cell lung cancer with pleural effusion. Patients are administered 4mg per kg body weight HPPH 24-48 hours before the surgery. Patients undergoing photodynamic therapy (PDT) are treated with light therapy with a fluence of 15-60 J/cm2 at 661nm. Fluence rate (mW/cm2) and cumulative fluence (J/cm2) is monitored at 7 different sites during the entire light treatment delivery. Isotropic detectors are used for in-vivo light dosimetry. The anisotropy of each isotropic detector was found to be within 15%. The mean fluence rate delivery and treatment time are recorded. A correlation between the treatment time and the treatment volume is established. The result can be used as a clinical guideline for future pleural PDT treatment.