Measurement of Central Corneal Thickness Using Ultrasound Pachymetry and Orbscan II in Normal Eyes
Mohammad Mehdi Sadoughi1, Bahram Einollahi1, Neda Einollahi1
1Ophthalmic Research Center, Shahid Beheshti University of Medical Sciences, Tehran, Iran.
Purpose:
To compare ultrasound pachymetry and Orbscan II for measurement of central corneal thickness (CCT) in normal eyes.
Methods:
The current study was performed at Labbafinejad Medical Center (LMC), Tehran, Iran. Three hundred eyes from 150 healthy individuals referred for keratorefractive surgery were assessed first by Orbscan II and then by ultrasound pachymetry, and CCT values were recorded and compared.
Results:
Overall, Orbscan II overestimated CCT as compared to ultrasound pachymetry by about 2.4% (mean values 547.6 ± 34.7 versus 534.8 ± 34.7, respectively, P < 0.001). The difference was more significant when CCT was less than 500 microns (mean values 493.2 ± 16.9 versus 479.9 ± 15.6, mean overestimation: 2.6%, P < 0.001). There was good linear correlation between the two methods (Pearson's correlation r = 0.968, P < 0.0001).
Conclusion:
Orbscan II has good correlation with ultrasound pachymetry for measurement of CCT in normal eyes; however Orbscan II should not be used to evaluate corneal thickness before keratorefractive surgeries, as it tends to overestimate corneal thickness and may result in undesirable, low residual stromal thickness.