A pilot study of the relationship between Doppler-estimated carotid and brachial artery flow and cardiac index
U Weber1, N J Glassford2, G M Eastwood2
1Department of Intensive Care, Austin Hospital, Melbourne, Victoria, Australia.
Insights
Doppler ultrasound measurements of peripheral arterial blood flow in humans showed a weak correlation with cardiac index. These findings suggest that peripheral blood flow cannot reliably estimate cardiac index non-invasively.
Area of Science:
- Cardiovascular Physiology
- Medical Imaging
- Diagnostic Ultrasound
Background:
- Cardiac index is a critical measure of heart function.
- Non-invasive methods for estimating cardiac index are highly sought after.
- Doppler ultrasound is widely used for assessing blood flow.
Purpose of the Study:
- To investigate the correlation between Doppler-estimated peripheral arterial blood flow (carotid and brachial arteries) and cardiac index in healthy human volunteers.
- To determine if peripheral arterial blood flow can serve as a non-invasive surrogate for cardiac index.
Main Methods:
- Doppler ultrasound was used to measure carotid and brachial artery blood flow in 11 human volunteers.
- Measurements were correlated with simultaneously measured cardiac index.
- Statistical analysis, including correlation coefficients (ρ), was employed.
Main Results:
- Moderate to good correlations were observed between right- and left-sided brachial (ρ = 0.45) and carotid (ρ = 0.567) arterial blood flows.
- Peripheral arterial blood flows (brachial and carotid) showed no or negative correlations with cardiac index (e.g., right brachial: ρ = -0.145; right carotid: ρ = -0.376).
Conclusions:
- Doppler-estimated peripheral arterial blood flows exhibit only a weak correlation with cardiac index in humans.
- Peripheral arterial blood flow measurements using Doppler ultrasound are not suitable for non-invasive estimation of cardiac index.
- Findings contrast with some previous studies, highlighting the limitations of this non-invasive approach.
Abstract:
We measured carotid and brachial artery blood flow by Doppler ultrasound in 11 human volunteers, and related these to cardiac index and to each other. The median (IQR [range]) carotid arterial blood flow was 0.334 (0.223-0.381 [0.052-0.563]) l.min(-1) on the right and 0.315 (0.223-0.369 [0.061-0.690]) l.min(-1) on the left. The brachial arterial blood flow was 0.049 (0.033-0.062 [0.015-0.204]) l.min(-1) on the right and 0.039 (0.027-0.054 [0.011-0.116]) on the left. Cardiac index was 3.2 (2.8-3.5 [1.9-5.4]) l.min(-1) .m(-2) . There was a moderate to good correlation between right-and left-sided flows (brachial: ρ = 0.45; carotid: ρ = 0.567). Brachial and carotid flow had no or a negative correlation with cardiac index (right brachial: ρ = -0.145, left brachial: ρ = -0.349; right carotid: ρ = -0.376, left carotid: ρ = -0.285). In contrast to some previous studies, we found that Doppler-estimated peripheral arterial blood flows only show a weak correlation with cardiac index and cannot be used to provide non-invasive estimates of cardiac index in man.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...