Murine Fetal Echocardiography
Published on: February 15, 2013
Safety Indices during Fetal Echocardiography at the Time of First-Trimester Scan Are Machine Dependent
Dragos Nemescu1, Anca Berescu2, Mircea Onofriescu1
1Department of Obstetrics and Gynecology, University of Medicine and Pharmacy "Gr.T.Popa", Iasi, Romania; "Cuza Voda" Obstetrics and Gynecology Hospital, Iasi, Romania.
Abstract:
The aim of our study was to evaluate the thermal index (TI) and mechanical index (MI), during the assessment of the fetal heart at the time of first-trimester scan, with different ultrasound machines. This was part of an observational study conducted in patients undergoing routine first-trimester screening. Cases were examined with Voluson E8 or 730Pro scanners using 2-8 MHz transabdominal probes. TI and MI were retrieved from the saved displays while in gray mode, color flow mapping and pulsed-wave (PW) Doppler examinations of the fetal heart and also from the ductus venosus (DV) assessment. We evaluated 552 fetal cardiac examinations, 303 (55%) performed with Voluson E8 and 249 (45%) with Voluson 730Pro ultrasound machines. The gray-scale exam of the heart and the PW Doppler DV assessment had TI values significantly lower for the Voluson E8 group (median, 0.04 vs. 0.2 and 0.1 vs. 0.2, respectively). The MI values from gray-scale and color flow mapping of the heart were significantly lower (median, 0.6 vs, 1.2 and 0.7 vs. 1) and for PW Doppler exam of the tricuspid flow were significantly higher (median 0.4 vs. 0.2) in the Voluson E8 group. The TI values from Doppler examinations of the heart, either color flow or PW imaging and MI values from DV assessment were not significantly different between the two groups. A different (newer) generation of ultrasound equipment provides lower or at least the same safety indices for most of the first-trimester heart examinations.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Ultrasonography
During an ultrasonography procedure, a handheld device called...