Related Experiment Video
Updated: Apr 11, 2026

MRM Microcoil Performance Calibration and Usage Demonstrated on Medicago truncatula Roots at 22 T
Published on: January 16, 2021
An improved shielded RF transmit coil for low-frequency NMR and MRI
V V Kuzmin1, C P Bidinosti2, M E Hayden3
1Laboratoire Kastler Brossel, ENS-PSL Research University, CNRS, UPMC-Sorbonne Universités, Collège de France, 24 rue Lhomond, F-75005 Paris, France.
Abstract:
We describe an actively shielded cylindrical RF transmit coil producing a highly uniform internal field (±0.5%) over an extended volume and a strongly suppressed (÷20) external field. Direct field mapping and experimental checks using in-situ NMR and MRI of polarised (3)He at low temperature demonstrate performance consistent with numerical field computations.
Related Concept Videos
Diamagnetic Shielding of Nuclei: Local Diamagnetic Current
Atomic Nuclei: Magnetic Resonance
Nuclear Magnetic Resonance (NMR): Overview
NMR spectroscopy generates a spectrum where the characteristic absorption frequencies of the sample are...
Magnetic Resonance Imaging
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
NMR Spectrometers: Resolution and Error Correction