Use of a prototype radioprotection cabin in vascular neuroradiology: Dosimetry and ergonomics
Joel Guersen1, Kaouthar Karmouche1, Jean Baptiste Moyon1
1Pôle d'imagerie et de radiologie interventionnelle, CHU de Clermont-Ferrand, 63003 Clermont-Ferrand, France.
Objectives:
The aim of this work was to compare the performance of a prototype radioprotection cabin in interventional neuroradiology, and to assess its suitability for routine use.
Materials And Methods:
The radioprotection cabin was a prototype derived from the CATHPAX AF(®) model. Three operators carried out 21 procedures (19 brain arteriographies and 2 embolizations) using the radioprotection cabin and not wearing the usual lead individual protection equipment (IPE), and 17 procedures (16 brain arteriographies and 1 embolization) wearing the standard lead IPE (vest, skirt, thyroid shield and goggles), and not using the radioprotection cabin. In all cases, thermoluminescent dosimeters (TLDs) were positioned at head, trunk, pelvic region, and upper and lower limbs to measure the dose equivalent for Hp(0.07) or Hp(3) that they received, attenuated by either the cabin or the lead IPE. Parallel to these dosimetric measurements, the ergonomics of the protection cabin were appraised by each radiologist after each procedure.
Results And Conclusion:
The cabin procured an overall reduction of 74% of the dose received on the whole body with Hp(0.07)=0.04 mSv ± 0.01 (CL=95%) against Hp(0.07)=0.12 mSv ± 0.04 (CL=95%) for the IPE. Body protection with the cabin was near complete, and close to 100% for the regions not protected by the usual IPE (e.g. the head). We also showed that design weaknesses noted by the operators that hampered procedures (light reflections, reduced hand mobility, awkward access to radioscopy pedal) could be remedied by maker's improvements to the prototype and minor changes in work habits.
More Related Videos
08:25Radiation Planning Assistant - A Streamlined, Fully Automated Radiotherapy Treatment Planning System
Published on: April 11, 2018
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Radiation: Applications
The average...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Biological Effects of Radiation
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...