[Patient-specific dose verification method using ArcCHECK for total marrow irradiation with intensity modulated arc
This study validates patient-specific dose verification for total marrow irradiation (TMI) using ArcCHECK with Volumetric Modulated Arc Therapy (VMAT) and Helical Tomotherapy (HT). ArcCHECK effectively verified VMAT plans, while a double-plan method proved suitable for Helical Tomotherapy dose verification.
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiotherapy Dosimetry
Background:
- Accurate dose verification is crucial for total marrow irradiation (TMI) to ensure treatment efficacy and minimize toxicity.
- Volumetric Modulated Arc Therapy (VMAT) and Helical Tomotherapy (HT) are advanced techniques used for TMI, each presenting unique dose verification challenges.
Purpose of the Study:
- To investigate and validate patient-specific dose verification methods using the ArcCHECK device for TMI delivered via VMAT and HT.
- To compare the efficacy of different ArcCHECK functionalities for VMAT and HT dose verification.
Main Methods:
- Dose verification was performed using ArcCHECK for head-neck, chest-abdomen, and pelvic regions in 8 patients undergoing TMI.
- For VMAT plans, the ArcCHECK merging function was utilized. For HT plans, a double-plan method (reference and delivery plans) was employed.
- Gamma analysis was used to assess dose passing rates for both VMAT and HT plans.
Main Results:
- VMAT plans achieved high γ-analysis passing rates: 98.9% ± 1.9% (head-neck), 98.4% ± 1.8% (chest-abdomen), and 97.4% ± 2.1% (pelvic).
- HT plans showed acceptable γ-analysis passing rates: 94.3% ± 1.5% (head-neck), 96.5% ± 1.2% (chest-abdomen), and 94.1% ± 1.9% (pelvic).
- The merging function of ArcCHECK proved effective for VMAT TMI dose verification.
Conclusions:
- Patient-specific dose verification using ArcCHECK with its merging function is a reliable method for VMAT-based TMI.
- The double-plan method is suitable for dose verification of HT TMI plans, especially when targets are distant from the setup center.
- These findings support the clinical implementation of ArcCHECK for robust TMI dose verification across different delivery techniques.