Related Experiment Video
Updated: Apr 11, 2026

Ion-Exchange Membranes for the Fabrication of Reverse Electrodialysis Device
Published on: July 20, 2021
Improvements in ion reflux: An electrodialytic eluent generation and suppression device for ion chromatography
Kyle Elkin1, John Riviello2, Hamish Small3
1Norwegian University of Life Sciences, N-1432 Ås, Norway.
Abstract:
This work describes a membrane based electrodialytic ion reflux device (IRD), which uses water as the pumped phase and integrates isocratic and gradient eluent generation and suppression. The current design incorporates several ion exchange membranes to create discrete chambers for suppression and eluent generation, while isolating the electrodes from the analytical stream. A small volume of recycled water can be used as the pumped phase while continuously refluxing the eluent ions. This current design permits electronically controlled eluent generation of at least 16.4μeq KOHmin(-1), while maintaining low suppressed background conductivity (<0.5μS/cm). The device was operated in gradient or isocratic mode continuously for up to 6 weeks. During this period, over 500 gradient and isocratic injections were performed, showing peak retention time precision below 1.5% RSD.
More Related Videos
Related Concept Videos
Ion-Exchange Chromatography
Ion Exchange
High-Performance Liquid Chromatography: Elution Process
Dialysis
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Capillary Electrophoresis: Instrumentation