Related Experiment Video
Updated: Apr 11, 2026

Author Spotlight: Innovating Thiol Quantification and Biomarker Detection for Oxidative Stress Research
Published on: June 28, 2024
Highly sensitive determination of reduced glutathione based on a cobalt nanoparticle implanted-modified indium tin
Tong Wang1, Wen Su, Zhengjun Xiao
1College of Chemistry, Beijing Normal University, Beijing 100875, PR China. hujingbo@bnu.edu.cn.
Abstract:
Cobalt nanoparticle modified indium tin oxide (CoNP/ITO) electrodes fabricated by ion implantation were applied for the detection of reduced glutathione (GSH). The CoNP/ITO electrode was characterized by scanning electron microscopy (SEM), energy dispersive X-ray (EDX) detector and X-ray photoelectron spectroscopy (XPS). The assay performance with regard to GSH were evaluated by cyclic voltammetry (CV) and chronoamperometry (I-t). The proposed sensor exhibited a much higher electrocatalytic activity toward the oxidation of GSH than the bare ITO electrode, with a detection limit of 5 nM. The CoNP/ITO electrode showed enhanced electrocatalytic properties, high sensitivity, good long-term stability and reproducibility as well as a rapid response to detect GSH. In addition, the CoNP/ITO electrode was also used to analyse the GSH concentration in eye drop samples, and the results were in good agreement with the labelled values.