Clinical evaluation of image quality and radiation dose reduction in upper abdominal computed tomography using
Atsushi Nakamoto1, Tonsok Kim1, Masatoshi Hori1
1Department of Radiology, Osaka University Graduate School of Medicine, 2-2, Yamadaoka, Osaka, Suita Zip: 565-0871, Japan.
Purpose:
To evaluate the image quality of upper abdominal CT images reconstructed with model-based iterative reconstruction (MBIR) in comparison with filtered back projection (FBP) and adaptive statistical iterative reconstruction (ASIR) on scans acquired with various radiation exposure dose protocols.
Materials And Methods:
This prospective study was approved by our institutional review board, and informed consent was obtained from all 90 patients who underwent both control-dose (CD) and reduced-dose (RD) CT of the upper abdomen (unenhanced: n=45, contrast-enhanced: n=45). The RD scan protocol was randomly selected from three protocols; Protocol A: 12.5% dose, Protocol B: 25% dose, Protocol C: 50% dose. Objective image noise, signal-to-noise (SNR) ratio for the liver parenchyma, visual image score and lesion conspicuity were compared among CD images of FBP and RD images of FBP, ASIR and MBIR.
Results:
RD images of MBIR yielded significantly lower objective image noise and higher SNR compared with RD images of FBP and ASIR for all protocols (P<.01) and CD images of FBP for Protocol C (P<.05). Although the subjective image quality of RD images of MBIR was almost acceptable for Protocol C, it was inferior to that of CD images of FBP for Protocols A and B (P<.0083). The conspicuity of the small lesions in RD images of MBIR tended to be superior to that in RD images of FBP and ASIR and inferior to that in CD images for Protocols A and B, although the differences were not significant (P>.0083).
Conclusion:
Although 12.5%-dose MBIR images (mean size-specific dose estimates [SSDE] of 1.13mGy) yielded objective image noise and SNR comparable to CD-FBP images, at least a 50% dose (mean SSDE of 4.63mGy) would be needed to maintain the subjective image quality and the lesion conspicuity.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
09:57Development and Evaluation of 3D-Printed Cardiovascular Phantoms for Interventional Planning and Training
Published on: January 18, 2021
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...