Comparison of dual rotating Scheimpflug-Placido, swept-source optical coherence tomography, and Placido-scanning-slit
Yong Woo Lee1, Chul Young Choi1, Geun Young Yoon1
1From Sungkyunkwan University School of Medicine (Lee, Choi), Kangbuk Samsung Medical Center, Department of Ophthalmology, Seoul, South Korea; the University of Rochester Medical Center School of Medicine (Yoon), Department of Ophthalmology, New York, New York, USA.
Purpose:
To compare measurements of corneal indices using dual rotating Scheimpflug-Placido, swept-source optical coherence tomography (OCT), and Placido-scanning-slit systems.
Setting:
Kangbuk Samsung Hospital, Seoul, South Korea.
Design:
Prospective evaluation of diagnostic tests.
Methods:
Corneal topography measurements were performed using dual rotating Scheimpflug-Placido (Galilei G2), swept-source OCT (Casia SS-1000), and Placido-scanning-slit (Orbscan IIz) systems. The intraclass correlation coefficients (ICCs) and Bland-Altman plots were used to evaluate the agreement between measurements.
Results:
Fifty post-refractive surgery eyes and 50 normal eyes were evaluated. The agreement in anterior keratometry and pachymetry between the 3 devices was high in both groups (ICC > 0.9). In both groups, the ICC values for posterior keratometry and eccentricity were high between the dual rotating Scheimpflug-Placido and swept-source OCT systems (ICC > 0.9) but not between the Placido-scanning-slit system and the other 2 systems. The Placido-scanning-slit system yielded much steeper values for posterior keratometry in both groups (P < .05). The ICC values for posterior corneal elevation were lower than 0.9 between all 3 devices. The dual rotating Scheimpflug-Placido and swept-source OCT systems showed relatively higher ICC values than the Placido-scanning-slit system in both groups. Maximum posterior elevations were highest with the Placido-scanning-slit system followed by the swept-source OCT system and then the dual rotating Scheimpflug-Placido system.
Conclusions:
Anterior keratometry obtained using 3 devices showed high degrees of agreement. Posterior keratometry and eccentricity showed greater agreement between the dual rotating Scheimpflug-Placido and swept-source OCT systems than with the Placido-scanning-slit system. The dual rotating Scheimpflug-Placido and swept-source OCT systems were equivalent in detecting the shape of the cornea and could be considered interchangeable.
More Related Videos
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...