Cardiac Magnetic Resonance Imaging at 7 Tesla
Published on: January 6, 2019
All-systolic non-ECG-gated myocardial perfusion MRI: Feasibility of multi-slice continuous first-pass imaging
Behzad Sharif1,2, Reza Arsanjani1,3,4, Rohan Dharmakumar1,2,4
1Biomedical Imaging Research Institute, Cedars-Sinai Medical Center, Los Angeles, California, USA.
Purpose:
To develop and test the feasibility of a new method for non-ECG-gated first-pass perfusion (FPP) cardiac MR capable of imaging multiple short-axis slices at the same systolic cardiac phase.
Methods:
A magnetization-driven pulse sequence was developed for non-ECG-gated FPP imaging without saturation-recovery preparation using continuous slice-interleaved radial sampling. The image reconstruction method, dubbed TRACE, used self-gating based on reconstruction of a real-time image-based navigator combined with reference-constrained compressed sensing. Data from ischemic animal studies (n = 5) was used in a simulation framework to evaluate temporal fidelity. Healthy subjects (n = 5) were studied using both the proposed approach and the conventional method to compare the myocardial contrast-to-noise ratio (CNR). Patients (n = 2) underwent adenosine stress studies using the proposed method.
Results:
Temporal fidelity of the developed method was shown to be sufficient at high heart-rates. The healthy volunteers studies demonstrated normal perfusion and no dark-rim artifacts. Compared with the conventional scheme, myocardial CNR for the proposed method was slightly higher (8.6 ± 0.6 versus 8.0 ± 0.7). Patient studies showed stress-induced perfusion defects consistent with invasive angiography.
Conclusion:
The presented methods and results demonstrate feasibility of the proposed approach for high-resolution non-ECG-gated FPP imaging of 3 myocardial slices at the same systolic phase, and indicate its potential for achieving desirable image quality (high CNR and no dark-rim artifacts).
More Related Videos
08:13In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
06:29Cardiac Magnetic Resonance for the Evaluation of Suspected Cardiac Thrombus: Conventional and Emerging Techniques
Published on: June 11, 2019
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System V: CT
Magnetic Resonance Imaging