Related Experiment Video
Updated: Apr 11, 2026

Rapid Nanoprobe Signal Enhancement by In Situ Gold Nanoparticle Synthesis
Published on: March 7, 2018
Signal-to-Noise Enhancement of a Nanospring Redox-Based Sensor by Lock-in Amplification
Pavel V Bakharev1, David N McIlroy2
1Department of Physics, University of Idaho, Moscow, ID 83844, USA. bakharevpavel@gmail.com.
Abstract:
A significant improvement of the response characteristics of a redox chemical gas sensor (chemiresistor) constructed with a single ZnO coated silica nanospring has been achieved with the technique of lock-in signal amplification. The comparison of DC and analog lock-in amplifier (LIA) AC measurements of the electrical sensor response to toluene vapor, at the ppm level, has been conducted. When operated in the DC detection mode, the sensor exhibits a relatively high sensitivity to the analyte vapor, as well as a low detection limit at the 10 ppm level. However, at 10 ppm the signal-to-noise ratio is 5 dB, which is less than desirable. When operated in the analog LIA mode, the signal-to-noise ratio at 10 ppm increases by 30 dB and extends the detection limit to the ppb range.
Related Concept Videos
Amplifying Signals via Enzymatic Cascade
Small-Signal Analysis of MOSFET Amplifiers