Effect of exposure parameters and voxel size on bone structure analysis in CBCT
R Pauwels1,2, T Faruangsaeng1, T Charoenkarn1
11 Department of Radiology, Faculty of Dentistry, Chulalongkorn University, Bangkok, Thailand.
Objective:
To evaluate the effect of exposure parameters and voxel size on bone structure analysis in dental CBCT.
Methods:
20 cylindrical bone samples underwent CBCT scanning (3D Accuitomo 170; J. Morita, Kyoto, Japan) using three combinations of tube voltage (kV) and tube current-exposure time product (mAs), corresponding with a CT dose index of 3.4 mGy: 90 kV and 62 mAs, 73 kV and 108.5 mAs, and 64 kV and 155 mAs. Images were reconstructed with a voxel size of 0.080 mm. In addition, the 90 kV scan was reconstructed at voxel sizes of 0.125, 0.160, 0.200, 0.250 and 0.300 mm. The following parameters were measured: bone surface (BS) and bone volume (BV) per total volume (TV), fractal dimension, connectivity density, anisotropy, trabecular thickness (Tb. Th.) and trabecular spacing (Tb. Sp.), structure model index (SMI), plateness, branches, junctions, branch length and triple points.
Results:
For most parameters, there was no significant effect of the kV value. For BV/TV, "90 kV" differed significantly from the other kV settings; for SMI, "64 vs 73 kV" was significant. For BS/TV, fractal dimension, connectivity density, branches, junctions and triple points values incrementally decreased at larger voxel sizes, whereas an increase was seen for Tb. Th., Tb. Sp., SMI and branch length. For anisotropy and plateness, no (or little) effect of voxel size was seen; for BV/TV, the effect was inconsistent.
Conclusions:
Most bone structure parameters are not affected by the kV if the radiation dose is constant. Parameters dealing with the trabecular structure are heavily affected by the voxel size.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...