Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
Sample stability for complete blood cell count using the Sysmex XN haematological analyser
Massimo Daves1, Elmar M Zagler1, Roberto Cemin2
1Clinical Pathology Laboratory, Hospital of Merano, Merano, Italy.
Haematological test results can be unreliable if samples are stored for over 3 hours, especially at elevated temperatures. Careful consideration of sample stability is crucial for accurate laboratory diagnostics.
Area of Science:
- Clinical Laboratory Science
- Hematology
- Biomedical Engineering
Background:
- Sample stability is critical for accurate hematology laboratory results.
- Investigating the reliability of Sysmex XN hematological testing is essential.
Purpose of the Study:
- To assess the impact of storage duration (up to 24 hours) and temperature (room temperature, 4°C, 37°C) on hematological test parameters.
- To determine the reliability of Sysmex XN analyzer in analyzing stored blood samples.
Main Methods:
- Whole blood samples from 16 healthy outpatients were analyzed immediately and after 3, 6, and 24 hours.
- Samples were stored in triplicate at room temperature, 4°C, and 37°C.
- Hematological parameters were measured using the Sysmex XN analyzer.
Main Results:
- Minimal bias was observed within 3 hours, except for RDW and PLT-I at 37°C.
- Significant bias in MCH, MCV, RBC, MCHC, and PLT-I emerged after 6 hours at various temperatures.
- After 24 hours, most parameters showed meaningful bias across all storage conditions, particularly at 37°C.
Conclusions:
- Hematological test results are significantly impacted by storage duration exceeding 3 hours.
- Storage temperature critically affects sample stability and test accuracy.
- Caution is advised when interpreting results from samples stored longer than 3 hours.