Discrepant INR values: a comparison between Manchester and Thrombotest reagents using capillary and venous samples
M Morrison1, A Caldwell, G McQuaker
1Department of Haematology, Monklands District General Hospital, Airdrie.
Clinical and Laboratory Haematology
|January 1, 1989
Summary
The International Normalized Ratio (INR) aims for consistent warfarin effect reporting. However, this study found discrepant INR values across different testing methods, potentially leading to incorrect warfarin dosing.
Area of Science:
- Clinical Pathology
- Pharmacology
- Medical Laboratory Science
Background:
- The International Sensitivity Index (ISI) calibrates thromboplastin reagents against WHO standards.
- The International Normalized Ratio (INR) aims to standardize reporting of warfarin's anticoagulant effect.
- Variations in laboratory techniques may impact INR accuracy and patient management.
Purpose of the Study:
- To evaluate the consistency of INR values across different common laboratory techniques.
- To identify potential discrepancies in warfarin effect reporting based on reagent and sample type.
- To assess the clinical implications of differing INR values on warfarin dosing.
Main Methods:
- Analysis of 128 warfarin patient samples.
- Comparison of INR values generated by four distinct methods: venous Thrombotest, capillary Thrombotest, venous Manchester reagent, and capillary Manchester reagent.
- Statistical analysis to identify significant differences between INR values.
Main Results:
- Significant discrepancies in INR values were observed across the four tested methods.
- Mean INR values from the Manchester venous reagent were consistently lower than those from the other three methods (P < 0.0001).
- This suggests a potential for under-dosing or over-dosing depending on the method used for dose adjustment.
Conclusions:
- Current laboratory techniques for determining INR values exhibit significant variability.
- The Manchester venous reagent appears to yield lower INR values compared to Thrombotest methods.
- These findings highlight a critical need for standardization to ensure accurate warfarin dosing and patient safety.