Optimization of Transesophageal Atrial Pacing to Assess Atrial Fibrillation Susceptibility in Mice
Published on: June 29, 2022
Effect of pulse width on transesophageal atrial pacing in dogs
Robert A Sanders1, Emily H Chapel2
1Department of Small Animal Clinical Sciences, College of Veterinary Medicine, Michigan State University, East Lansing, MI, USA.
Objective:
To determine the impact of stimulus pulse width (PW) on pacing threshold (PT), zone of capture (ZOC) and extraneous muscular stimulation (EMS).
Study Design:
Experimental trial in client-owned dogs.
Animals:
Seventeen dogs, median weight 16.1 kg (interquartile range: 11.4-21.5).
Methods:
Transesophageal atrial pacing (TAP) involved a 6 Fr pacing catheter inserted trans-orally into the esophagus to a position aboral to the heart in anesthetized dogs. The catheter was slowly withdrawn until atrial pacing was noted on an electrocardiogram. The catheter was withdrawn in 1 cm increments until TAP could not be achieved. PTs were recorded at each pacing site using PWs of 10.0, 5.0, 2.0 and 1.8 ms, always in that order.
Results:
The overall lowest mean PTs for all dogs were 6 ± 3 mA, 9 ± 4 mA, 11 ± 5 mA and 13 ± 5 mA at PWs of 10.0, 5.0, 2.0 and 1.8 ms, respectively. A significant decrease in overall minimum PT was noted using a PW of 10.0 ms compared with either 2.0 or 1.8 ms (p = 0.043 and p = 0.001, respectively) and pacing using 5.0 ms compared with 1.8 ms (p = 0.028). A significant increase in ZOC was noted using a PW of 10.0 ms compared with PWs of 5.0, 2.0 and 1.8 ms (p = 0.0047, p = 0.0006 and p = 0.0003, respectively), using a PW of 5.0 ms compared with PWs of 2.0 and 1.8 ms (p = 0.0011 and p = 0.0003, respectively) and using a PW of 2.0 compared with one of 1.8 ms (p = 0.0084).
Conclusions And Clinical Relevance:
Use of 10.0 or 5.0 ms PW to perform TAP minimized the power required to pace the atria, while a PW of 10.0 ms maximized the size of the ZOC.