Efficiency of a mechanical device in controlling tracheal cuff pressure in intubated critically ill patients: a
Saad Nseir1, Andrey Rodriguez, Paula Saludes
1Critical Care Center, Hospital de Sabadell, Universidad Autonoma de Barcelona, CIBER enfermedades respiratorias, Corporació Sanitaria Parc Tauli, Parc Tauli 1, 08208, Sabadell, Spain, s-nseir@chru-lille.fr.
Background:
Cuff pressure (P cuff) control is mandatory to avoid leakage of oral secretions passing the tracheal tube and tracheal ischemia. The aim of the present trial was to determine the efficacy of a mechanical device (PressureEasy®) in the continuous control of P cuff in patients intubated with polyvinyl chloride (PVC)-cuffed tracheal tubes, compared with routine care using a manometer.
Methods:
This is a prospective, randomized, controlled, cross-over study. All patients requiring intubation with a predicted duration of mechanical ventilation ≥48 h were eligible. Eighteen patients randomly received continuous control of P cuff with PressureEasy® device for 24 h, followed by discontinuous control (every 4 h) with a manual manometer for 24 h, or vice versa. P cuff and airway pressure were continuously recorded. P cuff target was 25 cmH2O during the two periods.
Results:
The percentage of time spent with P cuff 20-30 cmH2O (median (IQR) 34 % (17-57) versus 50 % (35-64), p = 0.184) and the percentage of time spent with P cuff <20 cmH2O (23 % (5-63) versus 43 % (16-60), p = 0.5) were similar during continuous control of P cuff and routine care, respectively. However, the percentage of time spent with P cuff >30 cmH2O was significantly higher during continuous control compared with routine care of tracheal cuff (26 % (14-39) versus 7 % (1-18), p = 0.002). No significant difference was found in P cuff (25 (18-28) versus 21 (18-26), p = 0.17), mean airway pressure (14 (10-17) versus 14 (11-16), p = 0.679), or coefficient of variation of P cuff (19 % (11-26) versus 20 % (11-25), p = 0.679) during continuous control compared with routine care of tracheal cuff, respectively.
Conclusions:
PressureEasy® did not demonstrate a better control of P cuff between 20 and 30 cmH2O, compared with routine care using a manometer. Moreover, the device use resulted in significantly higher time spent with overinflation of tracheal cuff, which might increase the risk for tracheal ischemic lesions.
Trial Registration:
Clinicaltrial.gov: NCT02109003.
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...