A Rapid and Chemical-free Hemoglobin Assay with Photothermal Angular Light Scattering
Published on: December 7, 2016
Clinical evaluation of a novel technology for non-invasive and continuous measurement of plasma haemoglobin
A J Broderick1, F Desmond1, G Leen2
1Department of Anaesthesia and Intensive Care Medicine, University College Cork, Cork, Ireland.
We undertook the first clinical evaluation of a novel, non-invasive device for the continuous measurement of plasma haemoglobin concentration in 25 patients undergoing elective cardiac surgery. At four pre-determined intervals, samples of blood were taken for plasma haemoglobin estimation on a blood gas analyser and a laboratory device and were compared with the plasma haemoglobin estimation on the novel device using the Bland-Altman method. The 95% limits of agreement for estimation of plasma haemoglobin concentration for the device vs. laboratory, the device vs. the blood gas analyser and the blood gas analyser vs. the laboratory were 101.3 g.l(-1) , 103.1 g.l(-1) and 14.5 g.l(-1) , respectively. The bias (mean difference) in each case was 27.4 g.l(-1) , 25.1 g.l(-1) and 2.4 g.l(-1) , respectively. We conclude that the novel device in its current form is not a suitable replacement for more invasive methods of determining plasma haemoglobin concentration in patients in the setting of cardiac surgery; however, lessons learnt from the study will help to improve the device's future performance.
We undertook the first clinical evaluation of a novel, non-invasive device for the continuous measurement of plasma haemoglobin concentration in 25 patients undergoing elective cardiac surgery. At four pre-determined intervals, samples of blood were taken for plasma haemoglobin estimation on a blood gas analyser and a laboratory device and were compared with the plasma haemoglobin estimation on the novel device using the Bland-Altman method. The 95% limits of agreement for estimation of plasma haemoglobin concentration for the device vs. laboratory, the device vs. the blood gas analyser and the blood gas analyser vs. the laboratory were 101.3 g.l(-1) , 103.1 g.l(-1) and 14.5 g.l(-1) , respectively. The bias (mean difference) in each case was 27.4 g.l(-1) , 25.1 g.l(-1) and 2.4 g.l(-1) , respectively. We conclude that the novel device in its current form is not a suitable replacement for more invasive methods of determining plasma haemoglobin concentration in patients in the setting of cardiac surgery; however, lessons learnt from the study will help to improve the device's future performance.
More Related Videos
07:24A Precision Medicine Tool for Measurement and Monitoring of Hemoglobin S in Sickle Cell Disease Patients Receiving Transfusion Therapy
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.