Clinical evaluation of a novel technology for non-invasive and continuous measurement of plasma haemoglobin

A J Broderick1, F Desmond1, G Leen2

  • 1Department of Anaesthesia and Intensive Care Medicine, University College Cork, Cork, Ireland.

Anaesthesia
|June 16, 2015
PubMed

We undertook the first clinical evaluation of a novel, non-invasive device for the continuous measurement of plasma haemoglobin concentration in 25 patients undergoing elective cardiac surgery. At four pre-determined intervals, samples of blood were taken for plasma haemoglobin estimation on a blood gas analyser and a laboratory device and were compared with the plasma haemoglobin estimation on the novel device using the Bland-Altman method. The 95% limits of agreement for estimation of plasma haemoglobin concentration for the device vs. laboratory, the device vs. the blood gas analyser and the blood gas analyser vs. the laboratory were 101.3 g.l(-1) , 103.1 g.l(-1) and 14.5 g.l(-1) , respectively. The bias (mean difference) in each case was 27.4 g.l(-1) , 25.1 g.l(-1) and 2.4 g.l(-1) , respectively. We conclude that the novel device in its current form is not a suitable replacement for more invasive methods of determining plasma haemoglobin concentration in patients in the setting of cardiac surgery; however, lessons learnt from the study will help to improve the device's future performance.

Related Concept Videos

Pulse rhythm01:30

Pulse rhythm

Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
1.7K
Special considerations while measuring oxygen saturation01:19

Special considerations while measuring oxygen saturation

Assessing respiratory rate concurrently with pulse measurement is fundamental to patient care, providing valuable insights into the patient's respiratory function. The normal breathing rate for an adult usually falls within a normal range of 12 to 20 breaths per minute. Abnormal respiratory rates can signal underlying health conditions or the need for immediate intervention.
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
1.1K
Assessing Blood pressure using a doppler ultrasound01:19

Assessing Blood pressure using a doppler ultrasound

To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
3.0K
Guidelines For Measuring Vital Signs01:19

Guidelines For Measuring Vital Signs

Following these guidelines can help nurses accurately measure vital signs, assess changes in patient conditions, and provide timely treatment when necessary. Adhering closely to the guidelines ensures the accuracy and reliability of the results.
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
3.4K