The appropriate whole-body index on which to base standardized uptake value in 2-deoxy-2-[(18)F]fludeoxyglucose PET
G Keramida1,2, J Hunter1, S Dizdarevic1,2
11 Clinical Imaging Sciences Centre, Brighton Sussex Medical School, University of Sussex, Brighton, UK.
Objective:
Tissue uptake of 2-deoxy-2-fluorine-18 fludeoxyglucose ((18)F-FDG) is routinely quantified as standardized uptake value (SUV), which in general is the fraction (F) of administered activity per millilitre of tissue multiplied by an index of body size, usually weight (W), i.e. F/ml × W = SUV or F/ml = SUV × (1/W). Other indices have been suggested as preferable to W, especially lean body mass (LBM) and body surface area (BSA). The second equation mentioned above shows that the reciprocal of the ideal index should correlate closely with F/ml and give a regression line through the origin. The purpose of this study was to determine which of these three indices best meets these criteria.
Methods:
Data were evaluated from 49 males and 51 females undergoing routine (18)F-FDG positron emission tomography/CT. A 3 cm diameter region of interest was drawn over the liver and F/ml recorded. LBM and BSA were estimated from height and weight.
Results:
Based on all patients, the reciprocals of the three indices gave similar correlation coefficients with F/ml, but only 1/LBM gave regressions close to the origin. Intercepts were significantly higher for females for 1/W and 1/BSA, consistent with females having more body fat, but there was no significant difference with 1/LBM.
Conclusion:
LBM is the best index on which to base SUV because adipose tissue accumulates less (18)F-FDG than other soft tissues.
Advances In Knowledge:
The value of this study lies in its use of a novel, more rational approach than previously to confirm that SUV should be based on LBM.
More Related Videos
07:28Studying Metabolic Brain Connectivity Using 2-Deoxy-2-[18F]Fluoro-D-Glucose Dynamic Positron Emission Tomography at the Single-subject Level
Published on: January 24, 2025
10:04Radionuclide-fluorescence Reporter Gene Imaging to Track Tumor Progression in Rodent Tumor Models
Published on: March 13, 2018
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
![Quantification of Atherosclerotic Plaque Activity and Vascular Inflammation using [18-F] Fluorodeoxyglucose Positron Emission Tomography/Computed Tomography FDG-PET/CT](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F3777.jpg&w=3840&q=50)