Related Experiment Video
Updated: Apr 9, 2026

Occlusion of the Great and Small Saphenous Vein Using Copolymeric Glue Based on N-Butyl Cyanoacrylate and Methacryloxy Sulfolane
Published on: December 9, 2022
Percutaneous Vascular Closure System Based on an Extravascular, Bioabsorbable Polyglycolic Plug (ExoSeal): Results
P Kamusella1, C Wissgott1, T Jahnke2
1Institute of Diagnostic and Interventional Radiology/Neuroradiology, Westküstenklinikum Heide - Academic Teaching Hospital of the Universities of Kiel, Lübeck, and Hamburg, Heide, Germany.
The ExoSeal polyglycolic acid plug system effectively achieved hemostasis in 93.9% of patients undergoing transfemoral interventions. This percutaneous closure system demonstrates safety and efficacy for 6F sheath sizes.
Area of Science:
- Vascular Surgery
- Interventional Cardiology
- Medical Devices
Background:
- Percutaneous transfemoral interventions are common but require effective hemostasis.
- Achieving rapid and safe closure of the arteriotomy site is crucial for patient recovery.
- Existing closure methods have varying success rates and complication profiles.
Purpose of the Study:
- To evaluate the effectiveness and safety of the ExoSeal percutaneous closure system.
- To assess the hemostasis success rate using a polyglycolic acid plug.
- To determine complication rates associated with the device.
Main Methods:
- Prospective study of 1000 patients undergoing 6F sheath interventions.
- Utilized the ExoSeal percutaneous vascular closure system.
- Follow-up included clinical assessment and color-coded duplex ultrasound at 24 hours and 6 weeks.
Main Results:
- Immediate hemostasis was achieved in 93.9% of patients (939/1000).
- Device positioning failure occurred in 6.1% due to anatomical or device issues, managed with manual compression.
- Minor complications occurred in 7.4%, including hematomas and pseudoaneurysms; one retroperitoneal bleed required transfusion.
Conclusions:
- The percutaneous ExoSeal closure system provides safe and effective hemostasis for 6F transfemoral punctures.
- The device is a viable option for achieving rapid arteriotomy closure.
- Further monitoring confirmed the system's safety profile.
More Related Videos
20:33A Case Series of Successful Abdominal Closure Utilizing a Novel Technique Combining a Mechanical Closure System with a Biologic Xenograft that Accelerates Wound Healing
Published on: July 4, 2019
07:51The C-seal: A Biofragmentable Drain Protecting the Stapled Colorectal Anastomosis from Leakage
Published on: November 4, 2010