Dynamic 1.5-T vs 3-T true fast imaging with steady-state precession (trueFISP)-MRI sequences for assessment of
K Sinko1, C Czerny2, R Jagsch3
11 Department of Cranio-Maxillofacial and Oral Surgery, Medical University Vienna, Vienna, Austria.
Objectives:
To compare the image quality of MRI scans produced with 1.5- and 3.0-T devices during functional test condition.
Methods:
65 MRI scans obtained with 1.5- (n = 43) or 3.0-T (n = 22) true fast imaging with steady-state precession (trueFISP) sequences from patients with a history of a cleft palate were evaluated. Two experts assessed the MRI scans, independently of each other, and blinded to the MRI technique used. Subjective ratings were entered on a five-point Likert scale. The median planes of three anatomical structures (velum, tongue and pharyngeal wall) were assessed in three functional states (at rest, during phonation of sustained "e" and during articulation of "kkk"). In addition, MRI scans taken during velopharyngeal closure were evaluated.
Results:
Under blinded conditions, both observers (radiologist and orthodontist) independently rated the quality of 1.5-T scans higher than that of 3.0 T. Statistical analysis of pooled data showed that the differences were highly significant (p < 0.009) in 4 out of 10 test conditions. The greatest differences in favour of 1.5 T were observed for MRI scans of the velum.
Conclusions:
1.5 T used with trueFISP may be preferable over 3.0-T trueFISP for the evaluation of the velopharyngeal structures in the clinical routine.
Related Concept Videos
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies IV: Magnetic Resonance Imaging
Magnetic Resonance Imaging
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):