Indurometer vs. Tonometer: Is the Indurometer Currently Able to Replace and Improve Upon the Tonometer?
S Vanderstelt1, O J Pallotta2, M McEwen2
11 Lymphedema Research Unit, Department of Surgery, Flinders University , Adelaide, Australia .
The Indurometer shows potential for assessing tissue fibrosis in lymphedema patients but requires modifications to reduce measurement variability compared to the Tonometer. Further research is needed for its clinical adoption.
Area of Science:
- Biomedical Engineering
- Medical Device Technology
- Lymphedema Research
Background:
- The Indurometer, developed by Flinders Biomedical Engineering, aims to quantify tissue fibrosis by measuring resistance to applied force.
- It is designed as a potential replacement for the existing Tissue Tonometer in clinical practice.
Purpose of the Study:
- To compare the current Indurometer model with the Tissue Tonometer.
- To assess variability and repeatability between the two devices.
- To evaluate how variability differs across International Society of Lymphology (ISL) stages.
Main Methods:
- A multicenter trial involving 180 participants with unilateral arm lymphedema.
- Three repeat measurements using both the Indurometer and Tonometer at the forearm and upper arm.
- Comparison of measurement variability across different ISL stages.
Main Results:
- The Indurometer yielded lower measurement values than the Tonometer.
- The Tonometer demonstrated less variability than the Indurometer at both measurement sites.
- Tonometer measurements were significantly less variable in ISL stages 1 and 2, with no significant difference in stage 3.
Conclusions:
- The Indurometer offers improved ease of use over the Tonometer.
- Modifications are necessary to enhance the Indurometer's measurement variability before it can replace the Tonometer for lymphedema assessment.
- A better understanding of the ISL classification system is crucial for accurate interpretation.
More Related Videos
07:21Author Spotlight: Advancing Agricultural Land Ecosystem Research with a Hydraulic Property Analyzer to Assess Soil Health
Published on: August 9, 2024
05:14Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Uncertainty in Measurement: Reading Instruments
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Pipe Flowrate Measurement
The orifice meter is a simple,...