Accuracy in Dental Medicine, A New Way to Measure Trueness and Precision
Published on: April 29, 2014
Comparison of physical quality assurance between Scanora 3D and 3D Accuitomo 80 dental CT scanners
Ahmed S Ali1,2, Dareen Fteita1, Jarmo Kulmala1
1Laboratory of Biophysics, Cell Biology and Anatomy, Institute of Biomedicine, University of Turku, Turku, Finland.
Background:
The use of cone beam computed tomography (CBCT) in dentistry has proven to be useful in the diagnosis and treatment planning of several oral and maxillofacial diseases. The quality of the resulting image is dictated by many factors related to the patient, unit, and operator.
Materials And Methods:
In this work, two dental CBCT units, namely Scanora 3D and 3D Accuitomo 80, were assessed and compared in terms of quantitative effective dose delivered to specific locations in a dosimetry phantom. Resolution and contrast were evaluated in only 3D Accuitomo 80 using special quality assurance phantoms.
Results:
Scanora 3D, with less radiation time, showed less dosing values compared to 3D Accuitomo 80 (mean 0.33 mSv, SD±0.16 vs. 0.18 mSv, SD±0.1). Using paired t-test, no significant difference was found in Accuitomo two scan sessions (p>0.05), while it was highly significant in Scanora (p>0.05). The modulation transfer function value (at 2 lp/mm), in both measurements, was found to be 4.4%. The contrast assessment of 3D Accuitomo 80 in the two measurements showed few differences, for example, the grayscale values were the same (SD=0) while the noise level was slightly different (SD=0 and 0.67, respectively).
Conclusions:
The radiation dose values in these two CBCT units are significantly less than those encountered in systemic CT scans. However, the dose seems to be affected more by changing the field of view rather than the voltage or amperage. The low doses were at the expense of the image quality produced, which was still acceptable. Although the spatial resolution and contrast were inferior to the medical images produced in systemic CT units, the present results recommend adopting CBCTs in maxillofacial imaging because of low radiation dose and adequate image quality.