Objective measurements of image quality in synchrotron radiation phase-contrast imaging versus digital mammography
Y Ruiz-Gonzalez1, M Perez-Diaz2, D Martínez-Aguila2
1Center for Studies on Electronics and Information Technologies, Central University of Las Villas, Villa Clara, Cuba. yuselyr@uclv.edu.cu.
Purpose:
Phase-contrast mammography with synchrotron radiation is an innovative X-ray imaging practice that improves the identification of breast lesions. Previous studies have proven the superiority of the mammography images taken in the phase-contrast modality using synchrotron radiation beams as compared with images taken in conventional mammography by subjective analyses. However, to our knowledge, no previous study has compared different acquisition systems in order to quantify this improvement by means of objective robust indicators. In this research, we intend to quantify the superiority of phase-contrast imaging by means of objective metrics of image quality.
Methods:
Images from the American College of Radiology Mammographic Accreditation Phantom were obtained at hospitals, in two digital mammography equipment and at the Elettra synchrotron radiation facility (Trieste, Italy), using free space propagation phase-contrast modality. Regions of interest were selected to analyze image quality at the fibers (phase object) and masses (area object) simulated on the phantom by means of the signal-to-noise ratio, the figure of merit, the contrast and the edge visibility.
Results:
The image contrast and edge visibility were significantly higher at the phase-contrast modality as compared with digital mammography equipment. The figure of merit using phase-contrast modality was higher for the fibers and comparable for the masses.
Conclusion:
The results showed an improvement of the contrast and edge visibility in phase-contrast images. These improvements may be important in the detection of small lesions and details.
More Related Videos
14:09Quantitative Optical Microscopy: Measurement of Cellular Biophysical Features with a Standard Optical Microscope
Published on: April 7, 2014
08:46Using Synchrotron Radiation Microtomography to Investigate Multi-scale Three-dimensional Microelectronic Packages
Published on: April 13, 2016
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies III: Computed Tomography