Related Experiment Video
Updated: Apr 8, 2026

Atomic Force Microscopy Cantilever-Based Nanoindentation: Mechanical Property Measurements at the Nanoscale in Air and Fluid
Published on: December 2, 2022
Atomic force microscope caliper for critical dimension measurements of micro and nanostructures through sidewall
Hui Xie1, Danish Hussain1, Feng Yang1
1State Key Laboratory of Robotics and Systems, Harbin Institute of Technology, 2 Yikuang, C1-507, HIT Science Park, 150080 Harbin, China.
Abstract:
A novel atomic force microscope (AFM) dual-probe caliper for critical dimension (CD) metrology has been developed. The caliper is equipped with two facing tilted optical fiber probes (OFPs) wherein each can be used independently to scan either sidewall of micro and nanostructures. The OFP tip with length up to 500 μm (aspect ratio 10:1, apex diameter ⩾10 nm) has unique features of scanning deep trenches and imaging sidewalls of relatively high steps with exclusive profiling possibilities. The caliper arms-OFPs can be accurately aligned with a well calibrated opening distance. The line width, line edge roughness, line width roughness, groove width and CD angles can be measured through serial scan of adjacent or opposite sidewalls with each probe. Capabilities of the presented AFM caliper have been validated through experimental CD measurement results of comb microstructures and AFM calibration grating TGZ3.