Effective Analysis of Human Exposure Conditions with Body-worn Dosimeters in the 2.4 GHz Band
Published on: May 2, 2018
Increased dose near the skin due to electromagnetic surface beacon transponder
Kang-Hyun Ahn1, Ryan Manger, Howard J Halpern
1St. Mary's Good Samaritan Inc.. 123@123.com.
Abstract:
The purpose of this study was to evaluate the increased dose near the skin from an electromagnetic surface beacon transponder, which is used for localization and tracking organ motion. The bolus effect due to the copper coil surface beacon was evaluated with radiographic film measurements and Monte Carlo simulations. Various beam incidence angles were evaluated for both 6 MV and 18 MV experimentally. We performed simulations using a general-purpose Monte Carlo code MCNPX (Monte Carlo N-Particle) to supplement the experimental data. We modeled the surface beacon geometry using the actual mass of the glass vial and copper coil placed in its L-shaped polyethylene terephthalate tubing casing. Film dosimetry measured factors of 2.2 and 3.0 enhancement in the surface dose for normally incident 6 MV and 18 MV beams, respectively. Although surface dose further increased with incidence angle, the relative contribution from the bolus effect was reduced at the oblique incidence. The enhancement factors were 1.5 and 1.8 for 6 MV and 18 MV, respectively, at an incidence angle of 60°. Monte Carlo simulation confirmed the experimental results and indicated that the epidermal skin dose can reach approximately 50% of the dose at dmax at normal incidence. The overall effect could be acceptable considering the skin dose enhancement is confined to a small area (~ 1 cm2), and can be further reduced by using an opposite beam technique. Further clinical studies are justified in order to study the dosimetric benefit versus possible cosmetic effects of the surface beacon. One such clinical situation would be intact breast radiation therapy, especially large-breasted women.
More Related Videos
09:38Determining Pain Detection and Tolerance Thresholds Using an Integrated, Multi-Modal Pain Task Battery
Published on: April 14, 2016
06:21Diffuse Optical Spectroscopy for the Quantitative Assessment of Acute Ionizing Radiation Induced Skin Toxicity Using a Mouse Model
Published on: May 27, 2016
Related Concept Videos
Biological Effects of Radiation
Mutations
Chromosomal Alterations Are Large-Scale Mutations
While point mutations are changes in a single nucleotide in...
Dual Nature of Electromagnetic (EM) Radiation
Wavelength is the distance between two consecutive peaks (the highest point) or troughs (the lowest point) in the wave. Frequency is the number of...
Absorption of Radiation
Transdermal Drug Delivery Systems