PP025. Urinary dipstick proteinuria testing - Does automated strip analysis offer an advantage over visual testing?
D A De Silva1, C Halstead2, A-M Côté3
1Department of Obstetrics and Gynaecology, University of British Columbia, Canada; Department of Pathology and Laboratory Medicine, Children's and Women's Health Centre of British Columbia and the University of British Columbia, Vancouver, Canada.
Pregnancy Hypertension
|June 25, 2015
Summary
Automated dipstick urinalysis for proteinuria in pregnancy is not more sensitive than visual testing. Both methods show excellent positive likelihood ratios but only fair to poor negative likelihood ratios.
Area of Science:
- Obstetrics and Gynecology
- Clinical Chemistry
- Diagnostic Accuracy Studies
Background:
- Visual urinary test strips are common for screening proteinuria in pregnancy due to cost and convenience.
- However, visual test strips have known limitations in sensitivity and specificity.
- Automated readers were recommended by NICE guidelines to improve proteinuria detection accuracy.
Purpose of the Study:
- To compare the diagnostic accuracy of automated versus visual urine dipstick testing.
- The study focused on detecting a random protein:creatinine ratio (PrCr) of ⩾30mg/mmol.
Main Methods:
- A prospective cohort study evaluated 233 urine samples from 160 pregnant women (inpatients and outpatients).
- Urine samples were split for point-of-care visual and automated dipstick testing and laboratory analysis of protein and creatinine.
- Diagnostic test properties were determined by comparing dipstick results to a laboratory-confirmed random PrCr ⩾30mg/mmol.
Main Results:
- Both visual and automated dipstick testing demonstrated low sensitivity (56.0% and 53.9%, respectively).
- Positive likelihood ratios (LR+) were high for both methods (15.0 and 24.6).
- Negative likelihood ratios (LR-) were fair to poor for both (0.46 and 0.47).
Conclusions:
- Automated dipstick urinalysis offers no significant sensitivity advantage over visual testing for proteinuria in a predominantly outpatient pregnancy setting.
- While both methods have strong positive predictive power, their negative predictive power is limited.
- The performance of automated systems may depend on the specific test strips and analysers employed.
More Related Videos
Related Concept Videos
Urine Studies I: Urinalysis
2.3K
Urinalysis is a widely used diagnostic test that analyzes urine's physical, chemical, and microscopic characteristics. Healthcare providers use it to detect and monitor various health conditions, including renal disease, urinary tract infections (UTIs), diabetes, and metabolic or systemic disorders.Components of UrinalysisUrinalysis consists of three primary components: physical, chemical, and microscopic examination. Each provides unique insights into the urine sample and, by extension, the...
2.3K
Automated Microbial Diagnostics
45
Automated diagnostic analyzers have transformed clinical microbiology by providing rapid and reliable methods for pathogen identification and antibiotic susceptibility testing. Among these systems, the Vitek 2 is widely used because it automates the traditionally labor-intensive processes of microbial identification (ID) and antibiotic susceptibility testing (AST), delivering standardized and timely results that are essential for effective patient care.Microbial Identification with ID CardsThe...