Clinical accuracy of inflationary oscillometry in pregnancy and pre-eclampsia: Omron-MIT Elite
Related Experiment Video
Updated: Apr 8, 2026

Non-invasive Optical Measurement of Cerebral Metabolism and Hemodynamics in Infants
Published on: March 14, 2013
Y Chung1, M C Brochut1, A de Greeff1
1Women's Health Academic Centre, King's Health Partners, King's College London, London SE1 7EH, United Kingdom.
Objective:
To evaluate the accuracy of the Omron MIT Elite in pregnancy and pre-eclampsia according to the British Hypertension Society protocol (BHS).
Design:
Prospective observational study.
Setting:
Antenatal clinics and wards at St. Thomas' Hospital (London, UK).
Population:
Forty-five pregnant women including 15 with pre-eclampsia.
Methods:
Nine sequential same arm blood pressure (BP) measurements were taken from each woman by trained observers, alternating between mercury sphygmomanometry and the test device.
Main Outcome Measures:
Grading criteria of the BHS protocol (A/B grade=pass; C/D=fail).
Results:
The Omron MIT Elite achieved a grade A/A in both pregnancy and pre-eclampsia. The mean difference (SD) between the mercury standard and the device in pregnancy was -1.1 (5.2)mmHg and 1.5 (4.8)mmHg for systolic and diastolic BP respectively compared to 0.2 (5.3)mmHg and 2.2 (5.5)mmHg in pre-eclampsia.
Conclusion:
The Omron MIT Elite can be recommended for use in pregnancy and pre-eclampsia according to the BHS protocol. To date, this is the most accurate automated BP device validated in pre-eclampsia.