Related Experiment Video
Updated: Apr 8, 2026

05:41
Using a Cell-Tracer Injection to Investigate the Origin of Neointima-Forming Cells in a Rat Saccular Side Wall Model
Published on: March 16, 2022
2.7K
Modified injector for optimal insertion of standard CTRs in lax zonules
Calin Petru Tataru1,2, Anca Cristina Dogaroiu2, Constantin Mihai2
1Department of Ophthalmology, "Carol Davila" University of Medicine and Pharmacy, Bucharest - Romania.
European Journal of Ophthalmology
|June 26, 2015
Summary
A modified capsular tension ring (CTR) injector improves surgical safety by allowing controlled insertion, reducing complications during cataract surgery for patients with lax zonules.
Area of Science:
- Ophthalmology
- Surgical Innovation
- Medical Device Design
Background:
- Capsular tension rings (CTRs) are crucial for stabilizing the lens capsule in cases of lax zonules during cataract surgery.
- Standard CTR insertion techniques carry risks due to uncontrolled manipulation of the ring's free end, potentially causing intraoperative incidents.
Purpose of the Study:
- To present a modified CTR injector design that enhances control during insertion.
- To eliminate the risks associated with uncontrolled maneuvering of the CTR free end.
Main Methods:
- A modified standard CTR injector featuring a metal anchoring element to secure the second eyelet of the CTR was developed.
- The modified device and insertion technique were evaluated in 50 adult patients undergoing cataract surgery with lax zonules.
Main Results:
- The improved injector significantly reduced injection duration and complexity.
- No intraoperative incidents or postoperative complications were observed using the modified device.
Conclusions:
- Fixating the second CTR eyelet enables controlled intraocular implantation, simplifying the surgical technique.
- This method reduces surgical complications linked to blind maneuvering of the CTR free end and minimizes disruption to remaining zonular fibers.